Professional XNR-400C Mfi Testing Equipment
Main Specification:
| Type | XNR-400C |
| Squeeze out the part of material | |
| Discharge port diameter | Φ2.095±0.005mm |
| The length of the discharge port | 8.000±0.025mm |
| Loaded barrel diameter | Φ9.550±0.025mm |
| Loaded barrel length | 152±0.1mm |
| The diameter of the piston rod head | 9.475±0.015mm |
| The length of the piston rod head | 6.350±0.100mm |
| 1, Standard Test load (eight) | |
| Ⅰ | 0.325kg = (piston rod + weight tray + insulation sleeve + weight body) |
| Ⅱ | 1.200 kg = (0.325 +2 No. 0.875 weight) |
| Ⅲ | 2.160 kg = (0.325 + 1.835 weight) |
| Ⅳ | 3.800 kg = (0.325 +4 3.475 weight) |
| Ⅴ | 5.000 kg = (0.325 +5 No. 4.675 weights) |
| Ⅵ | 10.000 kg = (0.325 +5 No. 4.675 weight + 5.000 weight) |
| Ⅶ | 12.500 kg = (5.000 +7 2.500 0.325 +5 4.675 weight +6 weights) |
| Ⅷ | 21.600 kg, = (0.325 +2 No. 0.875 + 1.835 ) |
| Measurement | MFR and MVR |
| Thermostat range | Room temperature - 450 ° C |
| Temperature accuracy | ±0.2°C |
| Temperature resolution | 0.1°C |
| Timing range | 0.1-999 s |
| Timing accuracy | 0.1s |
| Cut material way | Automatic or manual cut material way |
| Display | Color touch screen |
| Can print the test report | |
| Working conditions | Ambient temperature of 10 ° C -40 ° C; Relative humidity of 30% -80% |
| Instrument Dimensions | 4500×400×610(L × W × H) |